2 - Installation
28
In example B, the FLT output of one unit is connected to the INH input of another unit. A fault condition
in one of the units will disable all of them without intervention either by the controller or external
circuitry. The controller can be made aware of the fault via a service request (SRQ) generated by the
Questionable Status summary bit.
Figure 2-6. FLT/INH Examples
Controller Connections
The ac source connects to a controller either through a GPIB or an RS-232 connector.
GPIB Connector
Each ac source has its own GPIB bus address. AC sources may be connected to the bus in series
configuration, star configuration, or a combination of the two. You may connect from 1 to 15 ac sources
to a controller GPIB interface.
NOTE: The ac source is shipped from the factory with its GPIB address set to 5. This address
can be changed as described in Chapter 4 of this guide.
FLT Output
INH Input
B) FLT Example with Multiple Units
FLT Output
INH Input
Switch
(Normally
Open)
INH Common
INH Input
A) INH Example with One Unit
+
-
+
-
I
FLT INH
. . . .
+
-
+
-
I
+
-
+
-
I
FLT INH
. . . .
+
-
+
-
I
+
-
+
-
I
NOTE: Connectors
are removable