10.0 CALIBRATION
ACTION DISPLAYS
Press the [Tare/Zero] key during the self-
test at power on. The scale will ask to
enter the password.
“ Pi n”
Default Password is 0000.
Enter “0” four times.
Password can be changed in technical
parameters.
Press [Print].
“ Pi n” “ ----”
Select the scale to be set up by using the
[Local/Remote] key.
Press the [Print] key to enter the
technical section.
“tEch” “ LocAL” “ “
“tEch” “rEmo tE” “ “
Use the [U.Wt.] to select the weighing unit
to be used for setting the calibration for
the scale. The arrow in the “Weight”
window will indicate the unit selected.
Press the [Print] key to continue.
“tEch” “ Uni t” “ “
You will enter the first parameter -
Calibration. Display will ask to unload any
weight on the platform.
Press the [Print] key to continue.
“Un L o Ad”
Enter the calibration weight to be loaded
and press [Print].
Load the calibration weight onto the scale
and press [Print] again.
The Calibration is complete, the scale will
run the self-test during which the weight
should be removed.
“SEL” “ 0010”
“Lo Ad”