6.3.2 Entering a known Unit Weight
If the unit weight is known it can be entered using the keypad.
Enter the value of the unit weight using the numeric keys
followed by pressing [U. Wt.] within few seconds while the
display is flashing. If no action is initiated within a few seconds,
the "Unit Weight” display will revert to the previous value,
otherwise it will show the new value that has been entered.
The sample is then added to the scale and the weight will be
displayed as well as the quantity based upon the unit weight.
When weighing in kilograms the unit weight is shown in grams.
When weighing in pounds the unit weight is shown in pounds.
After the unit weight has been determined or entered, the scale
can be used for parts counting. The scale may be tared to
account for the net weight as discussed in an earlier section.
After the scale is tared, the items to be counted are added and
the "Count" display will show the number of items computed
using the weight and the unit weight.
It is possible to increase the accuracy of the unit weight at any
time during the counting process by manually entering the
sample quantity and then pressing [Smpl]. Ensure that the
quantity displayed matches the quantity on the scale before
pressing the key. The unit weight will be adjusted based upon a
larger sample quantity providing greater accuracy when
counting larger sample sizes.
6.3.3 Automatic update of unit weight
The scales will automatically update the unit weight when a
sample less than the initial sample count is added. A beep will
be heard when the value is updated. Check the quantity is
correct when the unit weight has been updated automatically.
To lock the unit weight and prevent Auto-update, press [U.Wt.].
This feature is turned off as soon as the number of items added
exceeds the count used as a sample.