6.3.4 Count pre-set or check-weighing
Check-weighing (or Count Pre-setting) is a procedure to cause an alarm to
sound when the net weight (or the number of items) on the scale meets or
exceeds a number stored in memory. The value to be stored is entered
from the keyboard.
SETTING OF PRESET LIMITS
It is possible to set a high and low limit for either counting or weighing
(using net weight). When the [Pst] key is pressed the user can select
either counting or weighing and then set the lower and upper limit.
For example:
ACTION DISPLAYS
Press [Pst]
“PST ” “ nEt “ “ “ if currently in the weighing mode
Press
[U.Wt/Units]
“PST ” “ cnt “ “ “ to toggle from weighing to counting
Press [Print]
“Hi cnt” “0.3 2 3 4” “ ” The current high count limit is
displayed, press [CE] to clear and change if needed.
Press [Print]
“lo cnt” “0.0 2 3 4” “ ” The current low count limit is
displayed, press [CE] to clear and change if needed.
Press [Print]
Return to weighing with limits set.
• Clearing of either or both the limits is allowed. Both limits cleared will
disable the Preset altogether.
• If Preset Weight was selected the first display would show “Hi net” and
“lo net”.
• The beeper will work as described in the Beeper Parameter.
© Adam Equipment Company 2007
17