Agilent Technologies 08495-90025 Video Games User Manual


 
10 8494/95/96G/H Operating And Service Manual
Installation
Installation
Initial Inspection Inspect the shipping container for damage. If the shipping container or
cushioning material is damaged, it should be kept until the contents of the
shipment have been checked for completeness and the instrument has been
checked mechanically and electrically. A procedure for checking electrical
performance is given under “Operator’s Check” on page 12 (also see
“Performance Tests” on page 14). If the contents of the shipment are
incomplete, if there is mechanical damage or defect, or if the instrument
does not pass the electrical performance test, notify the nearest Agilent
Technologies office. If the shipping container is damaged, or the cushioning
material shows signs of stress, notify the carrier as well as the Agilent office.
Keep the shipping materials for the carrier's inspection. The office will
arrange for repair or replacement without waiting for claim settlement.
NOTE Containers and materials identical to those used in factory packaging are
available through Agilent Technologies offices. If the instrument is being
returned to Agilent for servicing, attach a tag indicating the type of service
required, return address, model number, and full serial number. Also, mark
the container FRAGILE to assure careful handling. In any correspondence,
refer to the instrument by model number and full serial number.
Mating Connectors Mating RF connectors used with the Option 001 must be type-N male
connectors, which comply with U.S. military standard MIL-C-39012 (See
Figure 7 on page 19). For Option 002, male SMA connectors must be used.
For Option 003, 7 mm mating connectors must be used (Figure 6 on page
18).
The solenoid drive cable connector plug is a Viking Industries, Inc., part
number TKP12-102-P-TS-100-AU (see Replaceable parts). The connector
plug and contacts without the cable can be ordered from a Viking distributor.
When ordering the complete cable assembly from Agilent, use the part
number listed in Table 5, “Replaceable Parts,” on page 15.
CAUTION When installing the instrument, make sure that the connectors do not support
weight or bear torque. The preferred procedure is to set up all equipment in
position before connecting the instrument.