Agilent Technologies 08495-90025 Video Games User Manual


 
8494/95/96G/H Operating And Service Manual 17
Service
4. Close the gap between coupling nut with the nut on the extended-sleeve
connector.
To disconnect:
1. Loosen the coupling nut of the connector showing the wider gold band.
CAUTION Part the connectors carefully to prevent striking the inner conductor contact.
To care for the device:
1. Keep contacting surfaces smooth and clean. Irregularities and foreign
particles can degrade electrical performance.
2. Protect the contacting surfaces when the connector is not in use by
leaving the coupling sleeve extended.
3. Use lintless material and/or firm-bristles brush such as a tooth brush for
cleaning. If a cleaning fluid is needed, use isopropyl alcohol.