Agilent Technologies 08495-90025 Video Games User Manual


 
6 8494/95/96G/H Operating And Service Manual
Specifications
Specifications
Frequency Range
and Attenuation
Attenuation Accuracy (±dB): (Referenced from 0 dB)
Instrument Frequency Range Attenuation
8494G dc to 4 GHz 0 dB to 11 dB in 1 dB steps
8494H dc to 18 GHz 0 dB to 11 dB in 1 dB steps
8495G dc to 4 GHz 0 dB to 70 dB in 10 dB steps
8495H dc to 18 GHz 0 dB to 70 dB in 10 dB steps
8496G dc to 4 GHz 0 dB to 110 dB in 10 dB steps
8496H dc to 18 GHz 0 dB to 110 dB in 10 dB steps
8494G/H 8495G/H
8496G/H
8494G 8494H 8495G 8495H 8496G 8496H
Attenuation Selection
(dB)
dc–4
GHz
dc–12.4
GHz
12.4–18
GHz
dc–4
GHz
dc–12.4
GHz
12.4–18
GHz
dc–4
GHz
dc–12.4
GHz
12.4–18
GHz
1 10 0.2 0.3 0.7 0.2 0.5 0.6 0.2 0.5 0.6
2 20 0.2 0.3 0.7 0.4 0.7 0.8 0.4 0.7 0.8
3 30 0.3 0.4 0.7 0.5 0.9 1.2 0.5 0.9 1.2
4 40 0.3 0.4 0.7 0.7 1.2 1.6 0.7 1.2 1.6
5 50 0.3 0.5 0.7 0.8 1.5 2.0 0.8 1.5 2.0
6 60 0.3 0.5 0.8 1.0 1.8 2.4 1.0 1.8 2.4
7 70 0.4 0.6 0.8 1.2 2.1 2.8 1.2 2.1 2.8
8 80 0.4 0.6 0.8 - - - 1.3 2.4 3.2
9 90 0.4 0.6 0.8 - - - 1.5 2.7 3.6
10 100 0.4 0.6 0.9 - - - 1.6 3.0 4.0
11 110 0.5 0.7 0.9 - - - 1.8 3.3 4.4