Calibration Procedure . . . continued
"CAL READING 1 OF 5" thru "CAL
READING 5 OF 5" is displayed; then
"CALIBRATION UPDATED." The
procedure is finalized and the unit
returns back to main menu.
☛
During the Cal Reading, the read head must
remain down and stable or an error message may
occur.
If "PLEASE WAIT XX (1-30) SECONDS" is
displayed during calibration, continue to hold read
head down until calibration readings are over. This
will only occur if the calibration procedure is
performed within 30 seconds of a previous
measurement.
If "READ ZERO REFLECTANCE" is displayed
after reading the white spot, refer to zero
reflectance procedure (next page). This will only
occur if the calibration values for zero reflectance
have drifted.
CAL READING
1 OF 5
CAL READING
2 OF 5
CAL READING
3 OF 5
CAL READING
4 OF 5
CAL READING
5 OF 5
CALIBRATION
UPDATED
6-6