© Adam Equipment Company 2011 Page
33
The Calibration is complete, the scale
will run the self-test during which the
weight should be removed.
If calibration is not possible an error
message will be shown
“- - - -” “ - - - -” “ - - - - ”
“SPAn” “ PASS” “ ”
“SPAn” “ FAiL” “ LoW ” or
“HiGH
If The Calibration is being done while in the TECH parameters (section 12) the linearity is
set at the same time as calibration by performing the calibration in 2 steps,
Display will ask to unload any weight on
the platform.
Press the [Print] key to continue.
“nO L o Ad”
Load the first calibration weight onto the
scale
If the mass is different than the value
shown press the [CE] key then enter
the calibration weight to be used .
Then press [Print] to continue or press
[Zero] to complete with only one mass
being used.
“LoAd 1 ” “ 0010” “ KiLoS”
Load the second calibration weight
onto the scale
If the mass is different than the value
shown press the [CE] key then enter
the calibration weight to be used .
Then press [Print].
“LoAd 2” “ 0030” “ KiLoS”
The Calibration is complete, the scale
will run the self-test during which the
weight should be removed.
“- - - -” “ - - - -” “ - - - - ”
“SPAn” “ PASS” “ ”
If calibration is not possible an
error message will be shown
“SPAn” “ FAiL” “ LoW ” or
“HiGH