6.2 MEMORY FUNCTIONS
The [M+] key will add the results of a weighing into memory, regardless of
whether the local or remote scale is being used for the weighing.
6.2.1 Manual accumulation
• The values (weight and count) shown on the display can be added to
the values in the accumulator by pressing the [M+] key. The "Weight"
display will show the total weight, the "Count" display will show the total
accumulated count and the "Unit Weight" display shows the number of
times items have been added to the accumulation memory. The values
will be displayed for 2 seconds before returning to normal.
• The scale must return to zero or a negative number before another
sample can be added to the memory.
• More products can then be added and the [M+] key pressed again.
This can continue for up to 99 entries or until the capacity of the
“Weight” display is exceeded.
• To observe the total value stored, press the [M+] key when the scale is
at zero. The total will be displayed for 2 seconds.
• To clear the memory, first press [M+] to recall the total from the
memory and then press [CE] to clear all values from the memory.
6.2.2 Automatic accumulated total
• The scale can be set to automatically accumulate total when a weight is
placed on the scale. This eliminates the need to press the [M+] key to
store values into the memory. However the [M+] key is still active and
can be pressed to store the values immediately. In this case, the
values will not be stored when the scale returns to zero.
• See Section 7 for enabling Automatic Accumulation