A SERVICE OF

logo

35
| Page © Adam Equipment Company 2009
% Solid
Atro Moisture
Time Left
(if time limit is set)
The Alternate displays in order are:
Mass
% Moisture
% Solid
% Atro Moisture
% Atro Solid
Time Remaining for test if time limit set
During the test the user can stop the test at any time by pressing the [Start]
key again. Otherwise it continues until the stop condition is met.
When the test has stopped the beeper will sound 3 short beeps (if enabled,
see sec. 10.1.2) and the display will show the end result:
This display shows the test was automatically stopped as stability was
reached in 12:44 minutes. If it had stopped at the end of the time period it
would have TSTOP on the top line and if manually stopped it would have
MSTOP.
SOLID
83.88%
ATRO MOIST
3.78%
TIME LEFT
14:21
12:44 ASTOP
76.47%