Agilent Technologies E3634A Video Gaming Accessories User Manual


 
Chapter 2 Initial Operation
Output Checkout
30
Current Output Checkout
The following steps check basic current functions with a short across the
power supply’s output.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the
OFF annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
8V* or 25V** annunciator turns on); and the knob is selected for voltage
control.
2 Connect a short across (+) and (-) output terminals with an insulated
test lead.
Use a wire size sufficient to handle the maximum current (See "Table 7-1 Wire
Rating" on page 153 in chapter 7).
3 Enable the output.
The
OFF annunciator turns off and the 8V* or 25V**, OVP, and OCP
annunciators are lit. The
CV or CC annunciator turns on depending on the
resistance of the test lead. The blinking digit can be adjusted by turning the
knob. Notice that the display is in the meter mode. ‘‘Meter mode’’ means that
the display shows the actual output voltage and current.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
Limit annunciator will be blinking). Adjust
the voltage limit to 1.0 volt to assure CC operation. The
CC annunciator will
turn on. To go back to normal mode, press the key again or let the display
time out for several seconds .
5 Set the knob to the current control to check that the front-panel
ammeter properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the meter
mode (the
Limit annunciator is off). Check that the ammeter responds to knob
control and the voltmeter indicates nearly zero (the voltmeter will show the
voltage drop caused by the test lead).
Power
Output
On/Off
Display
Limit
Display
Limit
Voltage
Current
*For Agilent E3633A Model **For Agilent E3634A Model