Agilent Technologies E3633A Video Gaming Accessories User Manual


 
Chapter 4 Remote Interface Reference
System-Related Commands
96
SYSTem:VERSion?
This command queries the power supply to determine the present SCPI
version. The returned value is of a string in the form YYYY.V where the ‘‘Y’s’’
represent the year of the version, and the ‘‘V’’ represents a version number for
that year (for example, 1996.0).
*IDN?
This query command reads the power supply’s identification string. The power
supply returns four fields separated by commas. The first field is the
manufacturer’s name, the second field is the model number, the third field is
not used (always ‘‘0’’), and the fourth field is a revision code which contains
three numbers. The first number is the firmware revision number for the main
power supply processor; the second is for the input/output processor; and the
third is for the front-panel processor.
The command returns a string with the following format (be sure to dimension
a string variable with at least 40 characters):
HEWLETT-PACKARD,E3633A or E3634A,0,X.X-X.X-X.X
*RST
This command resets the power supply to its power-on state as follows:
Command E3633A state E3634A state
CURR
20 A 7 A
CURR:STEP
0.32 mA (typical value) 0.13 mA (typical value)
CURR:TRIG
20 A 7 A
CURR:PROT
22.0 A 7.5 A
CURR:PROT:STAT
ON ON
DISP
ON ON
OUTP
OFF OFF
OUTP:REL
OFF OFF
TRIG:DEL
00
TRIG:SOUR
BUS BUS
VOLT
0 V 0 V
VOLT:STEP
0.36 mV (typical value) 0.95 mV (typical value)
VOLT:TRIG
0 V 0 V
VOLT:PROT
22.0 V 55.0 V
VOLT:PROT:STAT
ON ON
VOLT:RANG
P8V (Low) P25V (Low)