Chapter 5 Error Messages
Calibration Errors
132
714 Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
715 Bad OCP cal data
The overcurrent protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
716 Bad DAC DNL error correction data
Invalid data measured during the calibration for DAC differential nonlinearity
error correction.
717 Cal OVP or OCP status enabled
Overvoltage protection status or overcurrent protection status is enabled. You
must set both overvoltage and overcurrent protection status to OFF before and
during the calibration.
718 Gain out of range for Gain Error Correction
The slope of the DAC gain is out of range. Hardware fails.
740 Cal checksum failed, secure state
741 Cal checksum failed, string data
742 Cal checksum failed, store/recall data in location 0
743 Cal checksum failed, store/recall data in location 1
744 Cal checksum failed, store/recall data in location 2
745 Cal checksum failed, store/recall data in location 3
746 Cal checksum failed, DAC cal constants
747 Cal checksum failed, readback cal constants
748 Cal checksum failed, GPIB address
749 Cal checksum failed, internal data
750 Cal checksum failed, DAC DNL error correction data