© Adam Equipment Company 2009
15
Calibration selectable values for all model:
•
Press the [Tare/
↵
↵↵
↵
] key, the display will show “Load” then add the
selected calibration mass value to the pan.
•
After the calibration mass is added, wait for the stable symbol to come
on and then press the [Tare/
↵
↵↵
↵
] key, the display will show “PASS”, and
then return to normal weighing after 3 seconds.
•
If it is outside of the previous calibration value by more than 5% ,it will
show “FAIL L” or “FAIL H” indicating the new calibration weight was too
low(FAIL L) or too high (FAIL H) in respect to the previous calibration
value.
LBK 3 LBK 6 LBK 15 LBK 30
1Kg/2kg/3kg 2Kg/4kg/6kg 5Kg/10kg/15kg 10Kg/20kg/30kg
2lb/4lb/6lb 4lb/8lb/12lb 10lb/20lb/25lb 20lb/40lb/60lb