Agilent Technologies Model 6641A: 3217A-00111 Video Gaming Accessories User Manual


 
Calibration 99
CAL:CURR:MON (Series 668xA only)
This command sets the power supply to the current monitor (I
MON
) calibration point that is then entered with
CAL:CURR[:DATA]. The output current must be calibrated after CAL:CURR:MON is performed.
Command Syntax CALibrate:CURRent:MONitor
Parameters <NRf+>
Examples CAL: CURR: MON CALibrate: CURRent: MONitor
Query Syntax
(None)
Related Commands CAL:CURR[:DATA] CAL:STAT
CAL:PASS
This command enters a new calibration password. The command is active only when the power supply is already in the
calibration mode. Unless it is changed subsequently to shipment, the password is the power supply’s four-digit model
number. If the password is set to 0, password protection is removed and CAL:STAT ON is unrestricted. A new password is
automatically stored in nonvolatile memory and does not have to be stored with the CAL:SAVE command.
Command Syntax
CALibrate:PASScode <NRf>
Parameters
<NRf>
Examples
CAL:PASS 6671 CAL:PASS 09.1993
Query Syntax
(None)
Related Commands
CAL:STAT
CAL:SAVE
This command saves any new calibration constants (after a current or voltage calibration procedure has been completed) in
nonvolatile memory.
Command Syntax:
CALibrate:SAVE
Parameters
(None)
Examples
CAL: SAVE
Query Syntax
(None)
Related Commands
CAL:CURR CAL:VOLT CAL:STAT
CAL:STAT
This command enables and disables the calibration mode. The calibration mode must be enabled before the power supply
will accept any other calibration commands. The first parameter specifies the enabled or disabled state. The second
parameter is the password. It is required if the calibration mode is being enabled and the existing password is not 0. If the
second parameter is not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query
statement returns only the state, not the password.
Whenever the calibration mode is changed from enabled to disabled, any new calibration constants are lost unless they have
been stored with CAL:SAVE.
Command Syntax:
CALibrate:STATe <bool> [,<NRf>]
Parameters
{0 | OFF} | {1 | ON} [,<NRf>]
*RST Value
OFF
Examples
CAL:STAT 1,6671 CAL:STAT OFF
Query Syntax
CALibrate:STATe?
Returned Parameters
{0 | 1 }
Related Commands
CAL:PASS CAL:SAVE