Agilent Technologies E4375A Video Gaming Accessories User Manual


 
6 - Language Dictionary
98
CF_NEG_DIDT_LE
The change in the magnitude of current during the standard measurement
interval that is negative, and the magnitude of the change is less than or equal to
the programmed limit.
CF_DVMAX_GE
The magnitude of the difference between the voltage and the maximum voltage
observed during the step that is greater than or equal to the programmed limit.
CF_DVMAX_LE
The magnitude of the difference between the voltage and the maximum voltage
observed during the step that is less than or equal to the programmed limit.
CF_DVMIN_GE
The magnitude of the difference between the voltage and the minimum voltage
observed during the step that is greater than or equal to the programmed limit.
CF_DVMIN_LE
The magnitude of the difference between the voltage and the minimum voltage
observed during the step that is less than or equal to the programmed limit.
CF_DIMAX_GE
The magnitude of the difference between the absolute value of current and the
maximum absolute value of current observed during the step that is greater than
or equal to the programmed limit.
CF_DIMAX_LE
The magnitude of the difference between the absolute value of current and the
maximum absolute value of current observed during the step that is less than or
equal to the programmed limit.
CF_DIMIN_GE
The magnitude of the difference between the absolute value of current and the
minimum absolute value of current observed during the step that is greater than
or equal to the programmed limit.
CF_DIMIN_LE
The magnitude of the difference between the absolute value of current and the
minimum absolute value of current observed during the step that is less than or
equal to the programmed limit.
time_test_type is one of:
CF_TEST_BEFORE Action is taken if the measurement test is true before the time
argument
CF_TEST_AT Action is taken if the measurement test is true at the time argument
CF_TEST_AFTER Action is taken if the measurement test is true after the time
argument
CF_TEST_BEFORE_STIMULUS Action is taken if the measurement test is true before the stimulus
is applied for this step. The outputs are in a reset condition during
this test. The time argument is ignored for this time_test_type.
time is in seconds relative to the beginning of the present step number
action is the action taken when the meas_test_type and time_test_type are both satisfied. The
action choices are:
CF_NEXT
Advance as soon as possible to the next step, short circuiting all remaining tests
in the present step.
CF_FAIL
Mark the cell as a failure and disconnect it. No further tests or steps are applied
to the cell.
NOTE: CF_ACR_LE and CF_ACR_GE can only be used with CF_ACR steps.
CF_DCR_LE and CF_DCR_GE can only be used with CF_DCR. steps.
Other meas_test_types cannot be used with CF_ACR or CF_DCR steps.
You cannot delete a single test. cfResetSeq deletes all sequence steps and tests, and should be
sent prior to defining a new set of sequence steps and tests. If a forming sequence is in progress
when cfSetSeqTest is given, an error is returned.
Tests are volatile and disappear when the ac power is turned off.