Agilent Technologies Serials 2933A-08856 to 08860 3025A-09316 and Video Gaming Accessories User Manual


 
Standard Commands for Programmable Instruments (SCPI)
93
*RST
Meaning and Type
Reset
Device State
Description
This command reset the power supply to a factory-defined state as defined below. *RST also
forces an
ABORt and an OUT:PROT:CLE command.
CURR[:LEV] [:IMM] * OUTP OFF VOLT[:LEV][:IMM]*
CURR[:LEV]:TRIG* OUTP:PROT:DEL* VOLT[:LEV][:TRIG]*
CURR:PROT:STAT OFF OUTP:REL OFF
INIT:CONT OFF OUTP:REL:POL NORM
*Model-dependent value. See ISET, DI,Y, and VSET in Table 3-7 of Section III..
Command Syntax *RST
*SAV
Meaning and Type
SAVE
Device State
Description
This command stores the present state of the power supply to the specified location in memory. Up to five states can be
stored. Because the memory locations are volatile, all saved states are lost when the power supply is turned off. When the
power supply is turned on, all memory locations are initialized to the factory defined state (see the
*RST command).
The following power supply parameters are stored by
*SAV:
CURR[:LEV][:IMM] OUTP[:STAT] OUTP:REL:POL
CURR:PROT:STAT OUTP:PROT:DEL VOLT[:LEV][:IMM]
OUTP:REL[:STAT]
*SAV causes a write cycle to nonvolatile memory. Non volatile memory has a finite maximum number of
write cycles. Programs that repeatedly cause write cycles to nonvolatile memory can eventually exceed the
maximum number of write cycles and cause the memory to fail.
Command Syntax*SAV <NRf>
Parameters 0 to 4
Example SAV 3