Agilent Technologies Agilent 6030A Video Gaming Accessories User Manual


 
100 VAC Input Power Option 100 77
On page 65 (Analog Programming, 2nd paragraph), change the second sentence to read: “Resistance of 0 to 3.33 K ohms
programs the output voltage from 0 to full scale. and a resistance of 0 to 4 K ohms programs the output current from 0 to
full scale.”.
On page 65 (Analog Programming, 4th paragraph), change the second sentence to read, “A voltage of 0 to 4.25 V programs
the output voltage from 0 to full scale and a voltage of 0 to 5 volts programs the output current from 0 to full scale.”.
On page 66 (Constant Voltage Output, Resistance Control), in the second sentence, change 4000 ohms to 3.33k ohms.
On page 66 (Constant Voltage Output, Voltage Control), change “0 to + 5 volts” to “0 to 4.25 volts”.
In Figures 3-10 and 3-11, where 5 Vdc, 5 E or 5 appear, change them to read +4.25 Vdc, 4.25E, or 4.25 respectfully.