Agilent Technologies Agilent 6030A Video Gaming Accessories User Manual


 
Operating Instructions
56
VSET?
and addressing the power supply to talk.
The power supply can be instructed to measure its actual output voltage by sending:
VOUT?
The results are placed on the GP-IB when the power supply is addressed to talk, in this format (using 20 as an example):
VOUT 20.000
NOTE The programming resolution (LSB) for the VSET and ISET commands are specified in Table 1-1. The
power supply will round off settings received to the nearest multiple of these values.
Current Setting. Current is programmed in either amps or milliamps using any of the following codes (the value 5 is used as
an example):
ISET 5
ISET 5 A
ISET 5 MA
The programmed current is the actual output if the power supply is in CC mode, or the current limit if the power supply is in
CV mode.
The current setting may be read by sending:
ISET?
and addressing the power supply to talk.
The power supply can be instructed to measure its actual output current by sending:
IOUT?
The results are placed on the GP-IB when the power supply is addressed to talk, in this format (using 5 as an example):
IOUT 5.000
OVP Measurement. OVP trip voltage is hardware set at the power supply front panel. The power supply can be instructed
to measure the OVP trip voltage by sending:
OVP?
The results are placed on the GP-IB when the power supply is addressed to talk, in this format (using 20 as an example):
OVP 20.000