Agilent Technologies 60502B Video Games User Manual


 
6
Table 60502-1 Specifications and Supplemental Characteristics (continued)
Source Turn-On Current Overshoot: Less than 10% of final value (in CC and CR modes when connected to power
supplies with voltage rise times of greater than 500
µ
s).
Programmable Short Circuit: 0.033
(0.002
typical)
Programmable Open Circuit: 20 k
(typical)
Drift Stability (over an 8 hour interval):
Current: ± 0.03% ± 10 mA
Voltage: ± 0.01% ± 10 mV
Reverse Current Capacity: 100 A when unit is on; 40 A when unit is off
Weight: 3.2 kg (7 lbs.)
Table 60502-2. Programming Ranges
Function Front Panel Front Panel HPSL Command Range of Values
Key Display (Short Form)
Constant Current
Set Range C:RNG value "CURR:RANG value"
Low Range
0 and
6 A
High Range
> 6 A and
60 A
Set Main Level CURR value "CURR value"
Low Range 0 to 6 A
High Range 0 to 60 A
Set Slew Rate
(shift)
C:SLW value "CURR:SLEW value"
Low Range
0.00001 to 0.5 (A/
µ
s)
High Range
0.001 to 5 (A/
µ
s)
Set Transient Level C:TLV value "CURR:TLEV value" same as main level
*Set Triggered Level "CURR:TRIG value" same as main level
Constant Resistance
Set Range R:RNG value "RES:RANG value"
Low Range
0 and
1
Middle Range
> 1
and
1 k
High Range
>1 k
and
10 k
Set Main Level RES value "RES value"
Low Range
0 to 1
Middle Range
1
to 1 k
High Range
10
to 10 k
Set Slew Rate
(shift)
Low Range V:SLW value "VOLT:SLEW value" same as voltage slew
Middle/High Range C:SLW value "CURR:SLEW value" same as current slew
Set Transient Level R:TLV value "RES:TLEV value" same as main level
*Set Triggered Level "RES:TRIG value" same as main level
Constant Voltage
Set Main Level VOLT value "VOLT value" 0 to 60 V
Set Slew Rate
(shift)
V:SLW value "VOLT:SLEW value"
0.001 to 0.5 (V/µs)
Set Transient Level V:TLV value "VOLT:TLEV value" same as main level
*Set Triggered Level "VOLT:TRIG value" same as main level