Agilent Technologies 60502B Video Games User Manual


 
5
Table 60502-1 Specifications and Supplemental Characteristics (continued)
Current Slew Rate:*
Rate # 60 A Range Step 6 A Range Step Transition Time
1 1 A/ms 0.1 A/s 8.0 ms
2 2.5 A/ms 0.25 A/s 3.2 ms
3 5 A/ms 0.5 A/s 1.6 ms
4 10 A/ms 1 A/ms
800
µ
s
5 25 A/ms 2.5 A/ms
320
µ
s
6 50 A/ms 5 A/ms
160
µ
s
7
0.1 A/
µ
s
10 A/ms
80
µ
s
8
0.25 A/
µ
s
25 A/ms
32
µ
s
9
0.5 A/
µ
s
50 A/ms
16
µ
s
10
1 A/
µ
s 0.1 A/
µ
s 12
µ
s
11
2.5 A/
µ
s 0.25 A/
µ
s 12
µ
s
12
5 A/
µ
s 0.5 A/
µ
s 12
µ
s
*AC performance specified from 3 to 60 V.
Voltage Slew Rate:
Rate # Voltage Range Step Transition Time*
1 1 V/ms 8.0 ms
2 2.5 V/ms 3.2 ms
3 5 V/ms 1.6 ms
4 10 V/ms
800
µ
s
5 25 V/ms
320
µ
s
6 50 V/ms
160
µ
s
7
0.1 V/
µ
s 85
µ
s
8
0.25 V/
µ
s 85
µ
S
9
0.5 V/
µ
s 85
µ
S
*Transition time based on low capacitance current source.
Resistance Slew Rate (1
range): Uses the value programmed for voltage slew rate.
Resistance Slew Rate (1 k and 10 k
ranges): Uses the value programmed for current slew rate.
Transient Current Overshoot (When programmed from 0A):
Range Transient Current Level Current Slew Rate Overshoot*
60 A 6-60 A All slew rates 0
3 A
1 A/µs to 5 A/µs
1%
3 A
1 A/µs to 0.5 A/µs
0
6 A 6 A All slew rates 0
3 A
0.25 A/µs to 0.5 A/µs
1%
3 A
0.1 A/ms to 0.1 A/µs
0
*Overshoot may be higher during the first five seconds of programming if unit has been operating at full
current. Overshoot values assume a total inductance of lµH, or less, in the load leads connected to the
D.U.T.