Agilent Technologies 60502B Video Games User Manual


 
8
Table 60502-4. Calibration Information
Ranges and
Calibration Points
Variables Variables
Value
Power Supply
Settings
Current
Shunt
High Current Range Hi_curr_rng 60 5 V/61 A 100 A
High Current Offset Hi_curr_offset 0.0282
Low Current Range Lo_curr_rng 6 5 V/10 A 15 A
Low Current Offset Lo_curr_offset 0.0197
Voltage Range N/A N/A 61 V/5 A N/A
Voltage Hi point Volt_hipt 60
Voltage Lo point Volt_lopt 2.7
Low Resistance Range Lo_res_rng 1 15 V/10.9 A 15 A
Low Resistance Hi point Lo_res_hipt 1
Low Resistance Lo point Lo_res_lopt 0.04
Middle Resistance Range Mid_res_rng 10 10.9 V/15 A 15 A
Middle Resistance Hi point Mid_res_hipt 30
Middle Resistance Lo point Mid_res_lopt 1
High Resistance Range Hi_res_rng 1001 60 V/6 A 15 A
High Resistance Hi point Hi_res_hipt 120
High Resistance Lo point Hi_res_lopt 12