PREPARATION PROCEDURE
Remove the protective cap of the ORP elec-
trode (MA924B/1).
DO NOT BE ALARMED IF SALT DEPOSITS
ARE PRESENT. This is normal with elec-
trodes. They will disappear when rinsed
with water.
During transport, tiny bubbles of air may
form inside the glass bulb affecting proper
functioning of the electrode. These bubbles
can be removed by “shaking down” the elec-
trode as you would do with a glass
thermometer.
If the bulb and/or junction is dry, soak the
electrode in MA9015 Storage Solution for at
least one hour.
If the filling solution (electrolyte) is more
FUSE REPLACEMENT
To change the fuse follow next steps:
Disconnect the power cord from the rear panel
of the instrument.
Pull out the fuse holder located near the power
cord connector.
Replace the fuse with a similar one.
Push the fuse holder with the fuse in the
appropriate place.
ELECTRODE CONDITIONING & MAITENANCE