ERROR 8 - LOW ABSORPTION
The measured absorption is too low. Possible reasons in-
clude:
1. empty reagent bottles;
2. clogged, pinched or broken tubes;
3. interrupted sample flow.
ERROR 9 - LOW BLANK READING
If the sample blank is highly colored or turbid, the light
level may be inadequate for a proper operation. This situ-
ation is indicated by the activation of the analyzer system
alarm relay and the appearance of “E09” on the display.
This condition can also be caused by failure of the LED
light source.
ERROR 10 - HIGH BLANK READING
Possible reason is optical LED brightness (DIAG 999) too
high.
Note Error codes E01 to E07 are displayed on the analyzer but
do not cause a system error condition.
Error codes E08 to E10 are not displayed
but cause the SYS.ERROR LED to turn on
and the corresponding relay to close.
To view these error codes, enter code
800.