SETTING OF PIN
• Display will show “F2” “Pin”. This parameter allows setting of a
new password number.
• Press [Tare]. The “WEIGHT” display will show "Pin 1"
• Enter the new password number. The “Unit Weight” display will
show dashes. Press [Tare].
• The “WEIGHT” display will change to "Pin 2 ", Enter the password
again and press [Tare].
• The display will show "donE" to show the new password has been
accepted and will return to the menu. Record the new password
in a secured place.
• To exit a parameter, press the [Zero] key.
8.0 CALIBRATION
• The CBC scales are calibrated using metric weights and CBCa
scales are calibrated using metric or pound weights depending on
the unit in use before calibration.
• The scale will display a value of the weight to be used for
calibration. You can enter a different value, if desired.
• For entering Calibration, see section 7.0 Parameters
• When “F1” “CAL” is displayed, press [Tare].
• The display will then show "unLoAd" to request all weight be
removed from the platform.
• Press the [Tare] key to set the zero point.