SETTING OF PIN
Display will show F2 Pin”. This parameter allows setting of a
new password number.
Press [Tare]. The “WEIGHT” display will show "Pin 1"
Enter the new password number. The “Unit Weight” display will
show dashes. Press [Tare].
The “WEIGHT” display will change to "Pin 2 ", Enter the password
again and press [Tare].
The display will show "donE" to show the new password has been
accepted and will return to the menu. Record the new password
in a secured place.
To exit a parameter, press the [Zero] key.
8.0 CALIBRATION
The CBC scales are calibrated using metric weights and CBCa
scales are calibrated using metric or pound weights depending on
the unit in use before calibration.
The scale will display a value of the weight to be used for
calibration. You can enter a different value, if desired.
For entering Calibration, see section 7.0 Parameters
When “F1” “CAL” is displayed, press [Tare].
The display will then show "unLoAd" to request all weight be
removed from the platform.
Press the [Tare] key to set the zero point.