Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function Range
1
Resolution Model 175 Model 177 Model 179
AC Volts
2,3
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2
DC Volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.15 % + 2 0.09 % + 2 0.09 % + 2
1000 V 1 V 0.15 % + 2 0.15 % + 2 0.15 % + 2
Continuity 600 1 Meter beeps at < 25 , beeper turns off at > 250 ; detects
opens or shorts of 250 µs or longer.
Ohms 600.0
6.000 k
60.00 k
600.0 k
6.000 M
50.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance 1000 nF
10.00 µF
100.0 µF
9999 µF
4
1 nF
0.01 µF
0.1 µF
1 µF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps
5
(True RMS)
(45 Hz to 1 kHz)
60.00 mA
400.0 mA
6.000 A
10.00 A
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 3 1.5 % + 3 1.5 % + 3
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of
3 at full scale up to 500 V, decreasing linearly to crest factor
1.5 at 1000 V.
3. For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3.
4. In the 9999
µ
F range for measurements to 1000
µ
F, the measurement accuracy is 1.2 % + 2 for all models.
5. Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.