For more specific cleaning procedures, refer to the electrode’s instruc-
tion manual.
IMPORTANT: After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water and recalibrate the
controller.
TROUBLESHOOTING
Evaluate your electrode performance based on the following.
Noise (Readings fluctuate up and down) could be due to -
clogged/dirty junction: Refer to the Cleaning Procedure above.
Dry Membrane/Junction: Soak in Storage Solution HI 70300
overnight. Check to make sure the installation is such as to create
a well for the electrode bulb to constantly remain moist.
Low Slope: Refer to the cleaning procedure above.
No Slope: - Check the electrode for cracks in glass stem or bulb
(replace the electrode if cracks are found).
- Make sure cable and connections are not dam-
aged nor lying in a pool of water or solution.
Slow Response/Excessive Drift: Soak the tip in Hanna Solu-
tion HI 7061 for 30 minutes, rinse thoroughly with distilled
water and then recalibrate the meter.
For ORP Electrodes: Polish the metal tip with a lightly abrasive
paper (paying attention not to scratch the surface) and wash
thoroughly with water.
Note: With industrial applications, it is always recommended to keep
at least one spare electrode handy. When anomalies are not
resolved with a simple maintenance, change the electrode
(and recalibrate the controller) to see if the problem is
alleviated.
SUGGESTED INSTALLATIONSSUGGESTED INSTALLATIONS
SUGGESTED INSTALLATIONSSUGGESTED INSTALLATIONS
SUGGESTED INSTALLATIONS
for pH/ORP ELECTRODESfor pH/ORP ELECTRODES
for pH/ORP ELECTRODESfor pH/ORP ELECTRODES
for pH/ORP ELECTRODES
The electrodes should be installed in such a away that they always lie
in the solution, whether in a well, tank or on the discharge pipe.
SHORT DISTANCE, INDOOR INSTALLATION
Due to the low currents involved, a very high grade of insulation is
required. A dry environment is needed in order to obtain a level of
insulation not lower than 10
12
.