OPERATIONAL GUIDE
ORP electrode CONNECTION
The BNC connector is protected
against vapors and humidity by a wa-
terproof sheath.
Slide the protective sheath down.
Connect the HI 3214P/2 ORP elec-
trode to the BNC connector and
then slide up. Make sure the con-
nector is completely covered.
Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with ORP elec-
trodes and they dissolve when rinsed with water.
HI 982401 ORP
Range ±999 mV
Resolution 1 mV
Accuracy (@20°C/68°F) ±5 mV
Typical EMC Deviation ±5 mV
Setpoint Selectable from 350 to 850 mV
Alarm LED blinks when the measured
ORP value is lower than the setpoint
Calibration Factory calibrated
Probe HI
3214P/2 interchangeable ORP electrode
(optional)
Casing IP54
Power Supply 12 Vdc adapter (included)
Dimensions 86 x 110 x 43 mm (3.4 x 4.3 x 1.7”)
Weight 150 g (5.3 oz.)
SPECIFICATIONS
3
P
R
When not in use, rinse the electrode with water and store
it with the protective cap.
If the electrode has been left dry, soak the tip in HI70300
storage solution for at least one hour to reactivate it.
Taking ORP Measurements
Turn the meter on.
Remove the protective cap from the
ORP electrode and immerse the tip
(4 cm/1½”) in the sample.
The LCD will show the ORP value
measured in mV unit. Allow the read-
ing to stabilize and HI982401 will
start continuous monitoring.
Note: to prevent damages to the meter,
remove the ORP electrode from
the sample before turning the
meter off.
Adjusting the Setpoint
To access the MEASURE/SET switch
and the setpoint adjustment trimmer,
unscrew the 4 screws on the meter
back and remove the rear panel.
Move the switch to the SET position
(left).