A SERVICE OF

logo

38394 V3.3 8/08
3
Specifications
Function Range Accuracy
400.0A (AC Amps 0.5 to 400.0A) ±(2% rdg + 5d)
(AC @ 50/60Hz)
AC/DC Current
600A ±(2% rdg + 8d)
AC Voltage 4.000V, 40.00V, 400.0V, 600 V ±(1.2% rdg + 5d)
(AC @ 50/60Hz)
400.0mV ±(0.5% rdg +2d) DC Voltage
4.000V, 40.00V, 400.0V, 600V ±(1.0% rdg + 2d)
400.0Ω, 4.000kΩ, 40.00kΩ, 400.0kΩ
±(1% rdg + 5d)
4.000MΩ
±(2% rdg + 2d)
Resistance
40.00MΩ
±(3.5% rdg + 5d)
Frequency ( 5V)
5.000Hz, 50.00Hz, .500.0Hz, 5.000kHz,
50.00kHz, 100.0kHz
±(1% rdg + 5d)
Continuity
Audible tone <10Ω approximately
Function
Maximum
Input
Function
Maximum
Input
V DC/AC 600V DC or
AC Peak
Frequency 250V DC or AC peak
Input Limits
A DC/AC 600A DC/AC,
fused
Resistance 400V DC or AC peak
(<10 seconds)
Conductor Size 1.18" (30mm) maximum
Battery type Two (2) 1.5V AA batteries
Range Selection Automatic ranging
Display 5000 Count LCD
Overload Indication "1" or “-1”
Low Battery Indication Battery icon
Environmental conditions Installation Category III (clamp); Category II (DMM),
Pollution degree 2; Altitude: 2000 meters, Indoor use
only
Operating Temperature/Humidity 32º to 122ºF (0
o
C to 50
o
C) / <80% RH
Storage Temperature/Humidity 14º to 140ºF (-10
o
C to 60
o
C) / <80% RH
Dimensions 7.0 x 2.5 x 1.3” (178 x 64 x 33mm)
Weight 8.3 oz. (230g)
Zero Adjust REL (Relative) push button
Diode Short/Open, Good/Defect Test
Sample Time 0.35 sec approx.
Input Impedance 10MΩ for ACV & DCV