7.0 TROUBLE SHOOTING
7.1 ERROR MESSAGES
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your Adam Equipment dealer for support.
ERROR
CODE
DESCRIPTION POSSIBLE CAUSES
Solutions
alue too high. Weight on the pan when
turning on.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Remove any weight from
the stainless steel pan.
Recalibrate.
Pan is not installed when
turning on.
Calibration not correct.
Damaged load cell.
Install pan and try power
on.
Recalibrate.
-----
correct when turning
the scale on.
Load cell damaged.
Electronics damaged.
Remove any weight form
the stainless steel pan and
try rebooting. Try
calibrating the machine.
Unit-E Unit error.
Low battery indicator.
Batteries may be flat. Change the batteries.
FAIL Calibration failure. Incorrect mass used to
calibrate. The user calibration
is not within 5% of factory
calibration.
Possible damaged to loadcell.
Try to recalibrate, check
that the weight used to
calibrate matches that of
the value of the scale.
Unstable – machine
cannot get a stable
reading.
Possible damage to the
mechanics / Loadcell.
Possible poor and drafty
location caused by
airconditioner or next to open
Replace load cell by Adam
Equipment dealer.
Make sure the balance is
on a flat surface and away
from vibration and air
movement.
No Power when
turning on.
Battery may be flat. AC
adapter may not be working.
Change the batteries. Or
use the AC adapter