General Safety Considerations
WARNING
This is a Safety Class 1 product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
product is likely to make the product dangerous. Intentional
interruption is prohibited.
WARNING
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
WARNING
For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 0.315A/250V for 100/120V
operation and 0.16A/250V for 220/240V operation). The use of other
fuses or materials is prohibited. Verify that the value of the line-
voltage fuse is correct.
• For 100/120V operation, use an IEC 127 5
×
20 mm, 0.315 A, 250 V, Agilent
part number 2110-0449.
• For 220/240V operation, use an IEC 127 5
×
20 mm, 0.16 A, 250 V, Agilent
Technologies part number 2110-0448.
CAUTION
Before switching on this instrument, make sure that the line voltage selector
switch is set to the line voltage of the power supply and the correct fuse is
installed. Assure the supply voltage is in the specified range.
CAUTION
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
CAUTION
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4
°
C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
CAUTION
Always use the three-prong ac power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
CAUTION
Do not
connect ac power until you have verified the line voltage is correct.
Damage to the equipment could result.