A SERVICE OF

logo

5.1.18 (I) RESTORE NVRAM
This command returns system boot-up parameters stored in non-volatile memory to their
factory default settings. This command is used in the manufacturing process when creating a
new scanner. It should be used in the field only when the NVRAM contents have been
destroyed by a hardware failure.
NOTE: This command does not restore tuned parameters particular to a specific
scanner.
When you press <I> , the prompt “ Are You Sure [Y/N] ? “ will appear. Hit <N> to escape to
the Diagnostic Tests Menu . Press <Y> to proceed.
Next , the following appears :
1 : CF1000
3 : CF300
4 : CF400/36
5 : CF500
Choice :
Enter the value corresponding to your scanner model.