5.1.8 MERGE AND SPLICE ADJUSTMENT (8)
This feature lets you precisely adjust the effective camera position to eliminate discontinuities
across camera boundaries. Merging is adjustment along the scan line. Splicing is adjustment
perpendicular to the scan line.
______________________________________________________________________
NOTE: Focus, magnification, merge, and splice functions are intertwined. If you adjust one,
you must test and, if necessary, adjust the other functions. This must be done in the following
order:
1. Focus
2. Magnification
3. Merge
4. Splice
______________________________________________________________________
When you press <8> to select this option, the prompt, “ Use Electronic Splice ? [N]”, appears.
If you reply with <N>, you will be able to adjust only the merge. If you reply with <Y> and
electronic splice is enabled in Global Parameters, both merge and splice can be adjusted. (To
enable electronic splice, set global parameter 2 to “ON.”)
The prompt, “Use Bar Pattern? [N],” the appears. For the purpose of this test, reply to the
prompt with <N> or just a <CR> .
----- Merge and Splice Adjustment -----
Boundary Indicator
0 1 2 3 4 5 6 7
Threshold 35 Range 00-63 Decimal
Merge 2F Range 00-FF Hex
Splice 0 Range -5 to +5
Motor Full Forward
Status Paused
Delay Counts 00 00 00 00 00 00 00 00
------------------------- MENU-------------------------
left arrow : Boundary Left right arrow : Boundary Right
up arrow : Threshold Up down arrow : Threshold Down
< or , : Increase Merge > or . : Decrease Merge
( or 9 : Increase Splice ) or 0 : Decrease Splice
space bar : Pause / Unpause
S : Select Scanning Speed R : Reverse Motor
F : Fast Paper Movement X : Exit
Figure 5-17 Merge and Splice Adjustment Menu